Multi-event November 2018 FDA Recall Drug by Bound Tree Medical, Llc

This Multi-event Class II drug recall was voluntarily initiated by Bound Tree Medical, Llc on November 2, 2018 for the product Drug. The FDA reported the reason for recall as labeling incorrect instructions. The product was distributed Nationwide and the recall is currently terminated.

Reported Recall Events

Recall Number: D-0386-2019

Reason for Recall
Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg.
Initiated
11-02-2018
Reported
01-30-2019
Quantity
339 kits (1 standard syringe / 1 IFU per kit)

Recall Profile & Regulatory Data

Class II - is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Bound Tree Medical, LLC
Nationwide in the USA
Curaplex Epi Kit NOT FOR IV USE, #8600-01120, Epi-Safe Kit 8600-01 120 contains:2 Safety-LOK Syringe, TB, 1cc w/ needle, 25 ga x5/8 , 2 Vials of epinephrine, 1mg/1mL, 4 Alcohol prep pads, 2 Adhesive bandages, 1 x 3 1 Epi Generic Injection. Rx Only. Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016

Batch or Lot Expiration Information

Lot# all lot numbers

Recall Number: D-0385-2019

Reason for Recall
Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg.
Initiated
11-02-2018
Reported
01-30-2019
Quantity
2,654 kits (6 syringes per kit / 1 IFU per kit)

Recall Profile & Regulatory Data

Class II - is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Bound Tree Medical, LLC
Nationwide in the USA
Curaplex Epi Safe Kit, 8600-01102 TRAINING ONLY, Rx Only, Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016

Batch or Lot Expiration Information

Lot# all lot numbers

Recall Number: D-0384-2019

Reason for Recall
Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg.
Initiated
11-02-2018
Reported
01-30-2019
Quantity
7,063 kits (1 syringe per kit / 1 IFU per kit)

Recall Profile & Regulatory Data

Class II - is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Bound Tree Medical, LLC
Nationwide in the USA
curaplex Epi Safe Kit, 8600-01101, contains:" 1ml Vial of Epinephrine " 1 Epi-Safe Syringe " 1 Safety Needle " 2 Alcohol Prep Pads " 1 Adhesive Dressing " 1 Insert, Rx Only. Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016

Batch or Lot Expiration Information

Lot# all lot numbers

Recall Number: D-0383-2019

Reason for Recall
Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg.
Initiated
11-02-2018
Reported
01-30-2019
Quantity
1,283 kits (8 syringes per kit / 3 IFU's per kit)

Recall Profile & Regulatory Data

Class II - is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Bound Tree Medical, LLC
Nationwide in the USA
curaplex Epi Safe Administration and Training Kits, #8600-01100, Contains: 2 Epi Safe Administration Kit (8600-01101), 1 Epi Safe Training Kit (8600-01102), Rx Only, Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016

Batch or Lot Expiration Information

Lot# all lot numbers