The following document was submitted to the FDA by the labeler of this product Air Liquide America L.p.. The document includes published materials associated whith this product with the essential scientific information about this product as well as other prescribing information. Product labels may durg indications and usage, generic names, contraindications, active ingredients, strength dosage, routes of administration, appearance, warnings, inactive ingredients, etc.

Oxygen Compressed Usp Label



WARNING: DO NOT HANDLE CYLINDER OR USE CONTENTS UNTIL 1) YOU HAVE READ AND UNDERSTAND THE MATERIAL SAFETY DATA SHEET FOR THIS PRODUCT. CALL 1-800-231-1366 FOR MORE COPIES. 2) YOU ARE TRAINED TO USE CYLINDER AND CONTENTS, INCLUDING EMERGENCY PROCEDURES.

OXYGEN COMPRESSED USP    OXYGEN 2

UN 1072   Rx ONLY

AIR LIQUIDE

AIR LIQUIDE AMERICA L.P.
P.O. BOX 460229
Houston, TX 77056

M62ALP-1   H08


Ez-Ox Label



EZ-OX™

IMPORTANT: · Before using secure the EZ-OX to keep it from falling or rolling over. · Inspect the regulator valve for traces of dirt, dust, oil or grease. Dirt and dust can be removed safely with a lint free cloth.

WARNING: If oil or grease is visible or suspected do not use the system.

OPERATING INSTRUCTIONS · With the regulator valve outlet facing away from you, activate the control knob clockwise to purge any undetectable debris from the regulator valve, then turn off the regulator valve · Attach the oxygen supply tubing to the regulator valve's barbed    outlet connection located inside the housing access hole on the right side. · Turn the graduated knob clockwise to the flow rate prescribed by the physician · When the pressure gauge shows 300 PSI, the cylinder should be exchanged for a full cylinder to ensure an adequate supply of oxygen · When finished using oxygen, turn the flow of oxygen to "OFF" by turning the knob counter-clockwise to the off position. If any hissing sound can be heard indicating a leak, do not force the knob counter-clockwise past the positive stop at the off position. Should a leak occur, remove the system from service and place it outdoors. Tag for defect and return it for repair. If the system is jarred, dropped or knocked over, it should be checked to ensure connections are secured, components are not damaged, and proper flow rate is maintained.

DO NOT ATTEMPT TO SERVICE OR REMOVE THE CYLINDER REGULATOR VALVE

Return to AIR LIQUIDE
AIR LIQUIDE

M63A-1 

D NOT REMOVE FROM HOSPITAL PROPERTY