FDA Label for Sterile Cord Blood Collection Unit (msc128d)

View Indications, Usage & Precautions

Sterile Cord Blood Collection Unit (msc128d) Product Label

The following document was submitted to the FDA by the labeler of this product Maco Productions. The document includes published materials associated whith this product with the essential scientific information about this product as well as other prescribing information. Product labels may durg indications and usage, generic names, contraindications, active ingredients, strength dosage, routes of administration, appearance, warnings, inactive ingredients, etc.

Maco Biotech Collection - Sterile Cord Blood Collection Unit (Msc128d)



The MSC128D Sterile Cord Blood Collection Unit consists of 300 mL collection bag containing 21 mL of Anticoagulant Citrate Phosphate Dextrose Solution USP (CPD), a 40 mL rinsing bag containing 8 mL of CPD and two 12 gauge needles with a protective shield (SECUVAM) for the used needle.

Sterile, non-pyrogenic fluid path. Sterilized by steam.

Rx only.

This product is not made with natural rubber latex.


Indications And Usage



MSC128D - For collection of up to 200 mL of umbilical cord blood. Use aseptic technique.


Warnings



  • Avoid contact with sharp objects.
  • Do not irradiate collected cord blood or components.
  • Making multiple punctures of the umbilical cord to increase collection volume may increase the risk of contamination.

Precautions



  • DO NOT use if the overwraps or bags show any signs of alteration.
  • This is a single use sterile kit: under no circumstances should it be reused.
  • For optimal cord blood quality it is recommended to maintain the cord blood at an ambient temperature (18-26°C / 64.4-78.8°F) or cold temperature (4-12°C / 39.2-53.6°F) and process within 48 hours of collection.1,2,3
  • If the collection volume is less than 60 mL, the cord blood should be processed within 24 hours of collection.


How Supplied



MSC128D Sterile Cord Blood Collection Unit contains:

  • 300 mL collection bag containing 21 mL CPD
  • 40 mL rinsing bag containing 8 mL CPD
  • 2-12 gauge needles with a protective shield (SECUVAM) for used needle
  • Graduation label as visual aid to approximate the volume.

Instructions For Usecarefully Read All The Instructions Before Use



Representative Product Drawing

I.       PREPARATION

  • Before use, verify the integrity of the collection unit (overwrap, outlet ports, tubing...).
  • Only contents inside the inner wrap are sterile and acceptable for use in a sterile field if pouch is unopened and undamaged.
  • II.    COLLECTION PROCEDURE

    • Close all clamps 3A, 3B and 3C (clamp 9 should remain open).
    • After clamping and sectioning of the umbilical cord, place the collection bag on a mixer and perform a venipuncture at lower end of the umbilical cord with first needle 1A.
    • Note:  Umbilical venipunctures can be done before or after the expulsion of the placenta according to your local procedures.

      3.  Open clamps 3A and 3C.

      4.  Stimulate cord blood flow by massaging the cord from the top downward to increase volume of collection.

      5.  When collection is completed, close clamps 3A and 3C.

      6.  Remove needle 1A and slide down SECUVAM 2A (needle protector) over needle 1A.

      7.  If necessary, perform a second umbilical venipuncture on the upper part of the umbilical cord (near the mother)

      [See Warnings]:

            a.  Open clamps 3B and 3C.

            b.  Perform venipuncture on the umbilical cord using other needle 1B.

            c.  When collection is completed, close clamps 3B and 3C.

            d.  Remove needle 1B and slide down SECUVAM 2B (needle protector) over needle 1B.


      III.    RINSING OF COLLECTION LINE

      • Open clamp 3C.
      • Snap the break-away cannula on top of rinsing bag 6.
      • Roll up bag 6 on itself to empty 8 mL of anticoagulant (CPD) in the tubing to push the cord blood remaining in the tubing into the collection bag.  Do not strip the tubing.
      • Close clamp 3C.
      • Proceed to section V END OF COLLECTION if collection volume is less than 150 mL.  For collection volumes greater than 150 mL, proceed to section IV ADDITION OF ANTICOAGULANT INTO THE COLLECTION BAG.
      • IV.    ADDITION OF ANTICOAGULANT INTO THE COLLECTION BAG

        Use this step only for collection volumes greater than 150 mL.

        • Seal the tubing above Y junction 5.
        • Move clamp 3C near to Y junction 5.
        • Using a blood stripper, strip the tubing from Y junction 5 towards collection bag 8 expelling the anticoagulant in the tubing into collection bag 8.
        • Repeat this operation 3 times.
        • Mix collection bag 8.
        • Close clamp 3C.
        • Proceed to section V END OF COLLECTION.
        • V.   END OF COLLECTION

          • Close clamp 9 to secure the collection.
          • Seal the tubing between clamp 9 and collection bag 8.
          • Fill in collection volume and CPD volume on the collection bag label.  The amount of CPD is as follows:
          •          a.  21 mL - if no additional CPD is added.

                     b.  29 mL - if additional 8 mL CPD from rinsing bag is added to the collection bag as directed in section IV.

            4.  For optimal cord blood quality, maintain the cord blood at an ambient temperature (18-26°C / 64.4-78.8°F) or cold temperature (4-12°C / 39.2-53.6°F).

            VI.    CONTACT INFORMATION

            Manufactured for United States of America (U.S.A.) and Canada by Maco Productions Polonia, Sp. z o.o.

            ul. Szwajcarska 22, 54-405 Wroclaw, POLAND

            +48 71 37 60 110

            For consumer information contact:

            • For the U.S.A.: Distributed by Uneklo USA, Inc. dba Macopharma USA
            •     3075 Breckinridge Boulevard, Suite 405, Duluth, GA 30096, USA

                  (770) 270-6867

              • For Canada: Distributed by Macopharma Canada
              •     1200 McGill College, Suite 1715, Montreal (Quebec) H3B 4G7

                    (514) 866-2222

                VII.   REFERENCES

                • Pope, B. Mitsakos, K., Bilgin, A., Hokin, B. and Grant, R.:  Predicting overall viability of cord blood harvests.  Transfusion 2012:52:1079-1085
                • Salge-Bartels U. Huber. M, Kleiner K, Volkers P, Seitz R, Heiden M:  Evaluation of Quality Parameters for Cord Blood Donations. Transfus Med Hemother 2009; 36:317-324
                • Prof. Dr. Kogler, Jose Carreras CBB, medical center, University of Duesseldorf:  General information on the institution and the production facility of the Jose Carreras Cord Blood Bank. 2014

Labels



Collection Bag Label

Overwrap Label

REF MSC128D