LOCAMETZ allows the direct preparation of gallium Ga 68 gozetotide injection with the eluate from one of the following generators (see below for specific instruction for use with each generator):
- Eckert & Ziegler GalliaPharm germanium-68/gallium-68 (68Ge/68Ga) generator
- IRE ELiT Galli Eo germanium-68/gallium-68 (68Ge/68Ga) generator
The instructions for use provided by the germanium-68/gallium-68 generator manufacturer should also be followed.
Prepare gallium Ga 68 gozetotide injection according to the following aseptic procedure:
a. Use suitable shielding to reduce radiation exposure.
b. Wear waterproof gloves.
c. Flip the cap off the LOCAMETZ vial and swab the septum with an appropriate antiseptic, then allow the septum to dry.
d. Pierce the LOCAMETZ vial septum with a sterile needle connected to a 0.2 micron sterile air venting filter to maintain atmospheric pressure within the vial during the reconstitution process.
e. Place the LOCAMETZ vial in a lead shield container.
f. Follow the generator specific procedures below. Schematic diagrams are provided in Figures 1 and 2.
Preparation with Eckert & Ziegler GalliaPharm Generator
1) Connect the male luer of the outlet line of the generator to a sterile elution needle (size 21G to 23G).
2) Connect the LOCAMETZ vial directly to the outlet line of the generator by pushing the elution needle through the rubber septum.
3) Elute directly from the generator into the LOCAMETZ vial.
4) Perform the elution manually or by means of a pump according to the generator instructions for use.
5) Reconstitute the lyophilized powder with 5 mL of eluate.
6) At the end of the elution, disconnect the LOCAMETZ vial from the generator by removing the elution needle and the vent needle with the 0.2 micron sterile air venting filter from the rubber septum. Then, invert LOCAMETZ vial once and place it upright.
7) Incubate the LOCAMETZ vial upright between 20°C to 30°C (68°F to 86°F) for at least 5 minutes without agitation or stirring.
8) After 5 minutes, assay the vial containing the gallium Ga 68 gozetotide injection for total radioactivity using a dose calibrator, calculate the radioactivity concentration and record the result.
9) After radiolabeling, gallium Ga 68 gozetotide injection may be diluted with Sterile Water for Injection, USP or 0.9% Sodium Chloride Injection, USP up to a final volume of 10 mL.
10) Perform quality controls according to the recommended methods in order to check compliance with the specifications [see Dosage and Administration (2.5)].
11) Store the LOCAMETZ vial containing the gallium Ga 68 gozetotide injection upright in a lead shield container below 30°C (86°F) until use.
12) After addition of gallium-68 chloride to the LOCAMETZ vial, use gallium Ga 68 gozetotide injection within 4 hours.
Preparation with IRE ELiT Galli Eo Generator
1) Connect the male luer of the outlet line of the generator to a sterile elution needle (size 21G to 23G).
2) Connect the LOCAMETZ vial directly to the outlet line of the generator by pushing the elution needle through the rubber septum.
3) Connect the LOCAMETZ vial through the vent needle with 0.2 micron sterile air venting filter to a vacuum vial (25 mL minimum volume) by means of a sterile needle (size 21G to 23G) or to a vacuum pump to start the elution.
4) Elute directly from the generator into the LOCAMETZ vial.
5) Reconstitute the lyophilized powder with 1.1 mL of eluate.
6) At the end of the elution, first withdraw the sterile needle from the vacuum vial or disconnect the vacuum pump in order to establish atmospheric pressure into the LOCAMETZ vial, then disconnect the vial from the generator by removing both the elution needle and the vent needle with the 0.2 micron sterile air venting filter needle from the rubber septum. Inversion of the LOCAMETZ vial is not needed.
7) Incubate the LOCAMETZ vial upright between 20°C to 30°C (68°F to 86°F) for at least 5 minutes without agitation or stirring.
8) After 5 minutes, assay the vial containing the gallium Ga 68 gozetotide injection for total radioactivity using a dose calibrator, calculate the radioactivity concentration and record the result.
9) After radiolabeling, gallium Ga 68 gozetotide injection may be diluted with Sterile Water for Injection, USP or 0.9% Sodium Chloride Injection, USP up to a final volume of 10 mL.
10) Perform quality controls according to the recommended methods in order to check compliance with the specifications [see Dosage and Administration (2.5)].
11) Store the LOCAMETZ vial containing the gallium Ga 68 gozetotide injection upright in a lead shield container below 30°C (86°F) until use.
12) After addition of gallium-68 chloride to the LOCAMETZ vial, use gallium Ga 68 gozetotide injection within 4 hours.
Figure 1. Preparation with Eckert & Ziegler GalliaPharm Generator
Figure 1. Reconstitution Procedure For Eckert & Ziegler Galliapharm Generator (Psma11 01)
Figure 2. Preparation with IRE ELiT Galli Eo Generator
Figure 2. Reconstitution Procedure For Ire Elit Galli Eo Generator (Psma11 02)