Description
The Ultra-Technekow™ V4 Generator is prepared with fission-produced molybdenum Mo-99 adsorbed onto alumina in a column shielded by lead, tungsten, or depleted uranium. The column assembly and shielding are encased in a plastic container that is covered with a plastic elution hood. The elution hood has an opening for the column assembly double inlet needles and an opening for the single outlet needle. The needles accommodate the sterile eluant vials and sterile evacuated collection vials. A sterile vial containing a bacteriostat is supplied with the generator for the customer to aseptically seal the outlet needle after each elution.
This terminally sterilized generator provides a closed system for the production of sterile metastable technetium Tc-99m, which is produced by the decay of molybdenum Mo-99. Incorporated between the column outlet and the collection vial is a sterile 0.22 micrometer filter. Sterile, non-pyrogenic isotonic solutions of Sodium Pertechnetate Tc 99m can be obtained conveniently by periodic aseptic elution of the generator. These solutions should be clear, colorless, and free from any particulate matter. The Sodium Pertechnetate
Tc 99m Injection is suitable for intravenous injection and direct instillation.
The carrier-free solution may be used as is, or diluted to the proper concentration. Over the life of the generator, an elution will contain an amount of technetium Tc-99m in direct proportion to the quantity of Mo-99 decay since the previous elution of the generator. The quantity of Tc-99m in the eluate is determined by quantity of Mo-99 on the column, and the elapsed time between elutions.
Each eluate of the generator should not contain more than the USP limit of 0.15 kilobecquerel molybdenum Mo-99 per megabecquerel technetium Tc-99m (0.15 microcurie Mo-99 per millicurie Tc-99m) per administered dose at the time of administration and an aluminum ion concentration of not more than 10 micrograms per milliliter of the generator eluate, both of which must be determined by the user before administration.
Since the eluate does not contain an antimicrobial agent, it should not be used after 12 hours from the time of generator elution.