Avance Nerve Graft Implant
FDA Label NDC 84545-270

Full FDA labeling including Indications, Dosage, Usage, and Precautions

Structured Product Label

The following Structured Product Label (SPL) was submitted to the FDA by Axogen Corporation for the product Avance Nerve Graft (NDC 84545-270). This document serves as the official prescribing information, containing essential scientific data and clinical materials required for healthcare providers and patients.

This specific version of the label includes detailed information regarding 1.1 sensory nerve discontinuity (≤25 mm), 1.2 sensory nerve discontinuity (>25 mm), 1.3 mixed and motor nerve discontinuity, 2.1 recommended dose, 2.2 preparation, 2.3 administration, 3 dosage forms and strengths, 4 contraindications, and other regulatory disclosures. Use the navigation below to review specific sections of the FDA submission.

Label Section Quick Index

1.1 Sensory Nerve Discontinuity (≤25 Mm)

AVANCE is indicated for the treatment of adult and pediatric patients aged one month and older with sensory nerve discontinuity ≤25 mm.

1.2 Sensory Nerve Discontinuity (>25 Mm)

AVANCE is indicated for the treatment of adult and pediatric patients aged one month and older with peripheral sensory nerve discontinuity >25 mm.

This indication is approved under accelerated approval based on improvement in static two-point discrimination (s2PD) at 12 months in sensory nerve gaps ≤ 25 mm, which provided empirical evidence to reasonably predict clinical benefit given similarities in pathophysiology and anticipated therapeutic effects. [see Clinical Studies (14)]. Continued approval for this indication may be contingent upon verification and description of clinical benefit in confirmatory clinical trials.

1.3 Mixed And Motor Nerve Discontinuity

AVANCE is indicated for the treatment of adult and pediatric patients aged one month and older with mixed and motor nerve discontinuity.

This indication is approved under accelerated approval based on improvement in s2PD at 12 months in sensory nerve gaps, which provided empirical evidence to reasonably predict clinical benefit given similarities in pathophysiology and anticipated therapeutic effects. [see Clinical Studies (14)]. Continued approval for this indication may be contingent upon verification and description of clinical benefit in a confirmatory clinical trial.

AVANCE is intended for surgical implantation.

  • The recommended number of AVANCE required is based on the nerve deficits identified for repair.
  • AVANCE may be trimmed to accommodate the size of the deficits. Each AVANCE is intended for implantation in a single patient only.
  • AVANCE is available in 16 size combinations of different lengths and diameters to fit the size and diameter of the nerve(s) to be treated. Healthcare professionals should carefully assess the length and diameter of the nerve deficit(s) to be treated to determine the most suitable quantity and sizes of AVANCE required for treatment. [see Dosage and Administration (2.3)]. 
  • Refer to Table 1 for the available sizes of AVANCE:

    Table 1. Available Sizes of AVANCE
    Length (mm)Diameter Range (mm)

    AVANCE is available in the lengths and diameters as shown in Table 1. The graft diameter is measured at each end of the graft, and variation along the length can occur in human tissue.

    Length Range (mm)

    15

    1 to 2

    15 to 21

    15

    2 to 3

    15 to 21

    15

    3 to 4

    15 to 21

    15

    4 to 5

    15 to 21

    30

    1 to 2

    30 to 36

    30

    2 to 3

    30 to 36

    30

    3 to 4

    30 to 36

    30

    4 to 5

    30 to 36

    50

    1 to 2

    50 to 56

    50

    2 to 3

    50 to 56

    50

    3 to 4

    50 to 56

    50

    4 to 5

    50 to 56

    70

    1 to 2

    70 to 76

    70

    2 to 3

    70 to 76

    70

    3 to 4

    70 to 76

    70

    4 to 5

    70 to 76

2.2 Preparation

Receipt and Storage of AVANCE:

AVANCE is shipped on dry ice in validated insulated shipping containers. AVANCE must remain frozen at ≤-40°C (≤-40°F) and be kept frozen until use. For a complete description of the AVANCE packaging configuration and storage conditions, [see How Supplied/Storage and Handling (16)].

AVANCE is to be prepared in an appropriate surgical environment.

Supplies:

The following supplies are not included but are needed for preparation:

  • Room temperature sterile normal saline or sterile Lactated Ringer’s solution (LRS).
  • Preparation Instructions:

    Following standard surgical site preparation for exposure, mobilization, and trimming of the peripheral nerve to be treated, determine the size and/or number of AVANCE that need to be thawed. [see Dosage and Administration (2.3)].

    The following instructions are for preparing one AVANCE. Two team members (one sterile field team member and one team member outside the sterile field) are required.

    If two or more AVANCE are required, prepare them individually. AVANCE thawing should take approximately 6 minutes.

    • 1.The team member outside the sterile field removes the desired AVANCE from the freezer and transports it to the operating room.
    • 2.The team member outside the sterile field removes the zip top bag and opens and removes the carton.
    • 3.The team member outside the sterile field peels open the seal of the outer foil-poly pouch and aseptically presents the inner Tyvek®-poly pouch to the sterile field team member for placement in the sterile field. The sterile field team member aseptically removes the inner Tyvek-poly pouch presented by the team member outside the sterile field from the inside of the outer foil-poly pouch and places it in the sterile field.
    • 4.In the sterile field, the sterile field team member peels open the inner Tyvek-poly pouch and removes the clamshell tray.
    • 5.The sterile field team member opens the clamshell tray and fills the pre-molded thawing reservoir with room temperature sterile normal saline or sterile LRS (Figure 1). Do not heat AVANCE or add heated saline or LRS to the graft.
    •  

      Figure 1 (Image 01)

      Figure 1 (Image 01)

      Figure 1: Thawing AVANCE with sterile saline or LRS.

      Figure 1: Thawing AVANCE with sterile saline or LRS.

      • 6.Do not implant a partially or fully frozen product. Allow AVANCE to thaw completely before use, which will take approximately 6 minutes. AVANCE must be thawed until it reaches room temperature without heating. Once thawed, AVANCE must be either implanted or discarded within 12 hours.

2.3 Administration

AVANCE is implanted in appropriate sterile conditions by a trained licensed physician. Follow the steps below for surgical implantation of AVANCE:

  • 1.Prior to implanting AVANCE, it is important to trim back the nerve to be repaired until healthy nerve tissue is reached. Healthy nerve tissue can be identified by pouching fascicles and punctate bleeding.
  • 2.After trimming, if applicable, place the extremity in extension and measure the deficit as well as the diameter of the nerve to be repaired.
  • 3.Select an AVANCE of an appropriate length that ensures there will be no tension on the graft when implanted and of a diameter that most closely approximates the nerve to be repaired. AVANCE may be trimmed to accommodate the size of the deficits.
  • 4.When handling AVANCE, only hold it by the outermost epineurium and avoid crimping or crushing the graft.
  • 5.Coaptation can be achieved by multiple methods. Surgical methods should be tailored to the type of injury and goals of reconstruction, as well as body region, nerve type, and size of discontinuity. One method is to use appropriately sized non-absorbable sutures such as 8-‑0 or 9-0, which should be placed only through the epineurium of the nerve. Either end of AVANCE can be coapted to the proximal nerve to be repaired.
  • 6.To bridge the defect, secure the sutures, ensuring that they are not overtightened. The nerve being repaired and AVANCE should be lightly touching, but they should not overlap.

3 Dosage Forms And Strengths

AVANCE will vary in color from white, off-white, pink, pale pink, and yellow to pale yellow. AVANCE is a sterile, surgical implant and is intended for administration to a single patient. The available sizes of AVANCE are provided in Table 1.

4 Contraindications

None.

5.1 Procedural Complications

Procedural complications have occurred with the AVANCE implantation procedure and have included pain, hyperesthesia, infection, implant site swelling, adhesions, hypertrophic scar formation, impaired motor or sensory function, bleeding, and neuroma formation [see Adverse Reactions (6)]. Additional procedural complications may include risk of bleeding and complications related to anesthesia. Monitor patients for procedural complications with AVANCE implantation and manage accordingly.

5.2 Transmission Of Infectious Diseases

Transmission of infectious diseases or agents may occur with AVANCE administration, as it is manufactured using tissues from human cadaveric donors. Donors are screened and tested for Human Immune-deficiency Virus 1 (HIV-1); Human Immune-deficiency Virus 2 (HIV-2); Hepatitis B Virus (HBV); Hepatitis C Virus (HCV); and Syphilis (Treponema pallidum). Additional testing for Human T-cell Leukemia-lymphoma Virus 1 (HTLV-1) and Human T-cell Leukemia-lymphoma Virus 2 (HTLV-2) may be performed as required by local regulatory authorities in US states. Only screening was performed for Creutzfeldt-Jakob disease (CJD). These measures do not eliminate the risk of transmitting these or other infectious diseases or agents.

All infections thought to be transmitted by AVANCE should be reported to Axogen Corporation at 1‑888-296-4361.

6.1 Clinical Trials Experience

Because clinical trials are conducted under widely varying conditions, adverse reaction rates observed in one clinical trial cannot be directly compared to rates in other clinical trials and may not reflect the rates observed in practice.

The safety database described in this section reflects exposure to AVANCE in the RECON Study. A total of 112 patients received AVANCE and 108 patients received NeuraGen® (bovine nerve cuff) and were followed for a duration of 12 months. [see Clinical Studies (14)].

A serious adverse reaction occurred in 1 patient, which was wound dehiscence.

Table 2 lists the most common adverse reactions that occurred in ≥ 2% of patients in the RECON Study.

Table 2. Adverse Reactions occurring in > 2% of Patients in the RECON Study

Adverse Reactions 

AVANCE
n = 112

n (%)

Bovine Nerve Cuff
n = 108

n (%)

  Implant site hyperesthesia 

3 (3)

5 (5)

  Procedural pain 

4 (4)

0

Other clinically significant adverse reactions that occurred in <2% of patients in AVANCE group include: implant site swelling (n=2), paraesthesia (n=2), hypertrophic scar (n=2), pyogenic granuloma (n=2), grade 3 neuroma (n=1), wound dehiscence (n=1), tendon adhesion (n=1), implant site nodule (n=1), osteomyelitis (n=1), and dermal cyst (n=1).

6.2 Postmarketing Experience

Adverse Reactions from Observational Studies

The safety of AVANCE was evaluated in an ongoing, multicenter, observational registry study in patients undergoing nerve repair. The registry assessed repairs using the Medical Research Council Classification (MRCC) across nerve types (sensory, motor, and mixed), gap lengths (5–210 mm), and body regions (upper extremity, head and neck, torso including breast, and lower extremity).

Adverse reactions reported in the study include neuroma, seromas, implant site infection, tissue necrosis, nerve regeneration failure, hyperesthesia, hypoesthesia, AVANCE implantation error, and hematoma.

Risk Summary

There are no data on the use of AVANCE in pregnant women. No animal reproductive and developmental toxicity studies have been conducted with AVANCE to assess whether it can cause harm to the mother or fetus when administered to a pregnant woman.

In the U.S. general population, the estimated background risk of major birth defects and miscarriage in clinically recognized pregnancies is 2% to 4% and 15% to 20%, respectively.

There is no data on the effect of AVANCE on human milk, the effect on the breastfed infant, or the effect on milk production.

8.4 Pediatric Use

The safety and effectiveness of AVANCE have been established in pediatric patients aged one month and older. The use of AVANCE in pediatric patients for the treatment of nerve discontinuity was supported by extrapolation of adult data from the RECON Study. [see Clinical Studies (14)].

8.5 Geriatric Use

There were 8 patients (4%) who were 65 years of age and older in the RECON Study. Clinical studies of AVANCE did not include sufficient numbers of patients aged 65 years and older to determine whether they respond differently from younger subjects.

11 Description

AVANCE (acellular nerve allograft-arwx) is a sterile, acellular nerve scaffold derived from human cadaveric peripheral nerve tissue for surgical implantation. The processing method maintains the three-dimensional scaffold of the native peripheral nerve, including the endoneurial tubes, perineurium, epineurium, and microvasculature of the extracellular matrix (ECM). The processing removes cellular and noncellular components, including cells, fat, blood, axonal debris, and glycosaminoglycans. AVANCE is saturated in Lactated Ringer’s Solution prior to packaging and storage.

12.1 Mechanism Of Action

AVANCE is an acellular nerve scaffold intended to restore nerve function in the distal tissue.

AVANCE provides structural and biochemical cues for axonal regeneration. The structural cue is provided by the intact endoneurial tubes. The biochemical cue is provided by the basement membrane proteins of the endoneurial tubes. The basement membrane of the endoneurial tubes includes many proteins, such as collagen, laminin, fibronectin, and proteoglycans. The laminin lining the endoneurial tubes of AVANCE has been shown in in vitro assays and animal studies to be bioactive by supporting Schwann cell migration, axon growth cone interactions, and neurite extension. Without the specific bioactive laminin composition, the neuroregenerative capacity is hindered. However, the exact mechanism of action is unknown.

12.2 Pharmacodynamics

The pharmacodynamic effects of AVANCE are not known.

12.3 Pharmacokinetics

The pharmacokinetic effects of AVANCE are not known.

13.1 Carcinogenesis, Mutagenesis, Impairment Of Fertility

Conventional carcinogenicity and reproductive and developmental toxicity studies have not been conducted. Results from an in vitro bacterial reverse mutation assay demonstrated that AVANCE extracts are non-mutagenic.

14 Clinical Studies

The efficacy of AVANCE was evaluated in a multicenter, prospective, randomized, patient- and evaluator-blinded study that compared implantation of AVANCE to implantation of NeuraGen nerve cuff (RECON; NCT01809002). The study enrolled adult patients with sustained injury to at least one sensory nerve, distal to the superficial palmar arch and proximal to the distal interphalangeal joint, and a nerve gap between 5 mm and 25 mm. Patients with nerve injuries longer than 24 weeks, injuries requiring replantation of the target digit, injuries with vascular damage resulting in inadequate perfusion despite repair, prior history of type 1 Diabetes Mellitus or type 2 Diabetes Mellitus requiring regular insulin therapy, and prior history of neuropathy were excluded.

A total of 220 patients were intraoperatively randomized in a 1:1 ratio to receive either AVANCE (n=112) or NeuraGen (n=108). Randomization was stratified based on nerve gap length (5–14 mm: Group A; 15–25 mm: Group B).

The population demographics were as follows: the median age was 37 years (range 18 to 69 years), and 155 patients (71%) were male. 174 patients (79%) were White, 33 patients (15%) were Black, 3 patients (1.4%) were Hispanic or Latino, 2 patients (0.9%) were Asian, 2 patients (0.9%) were Native Hawaiian or Pacific Islander, and 6 patients (2.7%) were of “other race”.

The primary efficacy outcome measure was improvement in static two-point discrimination (s2PD) in the target digit, with a non-inferiority test (non-inferiority margin of -2 mm) between the groups at 12 months following reconstruction.

The efficacy results are summarized in Table 3 below.

Table 3: Efficacy Results of RECON Study
a Estimated using repeated measures Analysis of Covariance (ANCOVA) model with fixed effects for repair type, visit, and gap length and a random effect for subject. Visits Pre-Op screening, Months 1, 3, 6, 9, and 12 were included.
CI=confidence interval; LS= least squares; mm=millimeter

Endpoints

AVANCE
(n = 112)

Nerve Cuff
(n = 108)

LS mean difference

Static two-point discrimination (s2PD) in the target digit at 12 monthsa

  LS mean

  (95% CI)

9.1 mm

(8.11, 10.04)

9.4 mm

(8.50, 10.30)

0.3 mm

(-0.98, 1.64)

16.1 How Supplied

AVANCE is shipped on dry ice and supplied frozen, as a single, intact graft of varying color from white, off-white, pink, pale pink, and yellow to pale yellow. Do not use if broken or otherwise compromised.

AVANCE is sterile packaged. The immediate container closure system consists of a clamshell (thermoformed tray) within two pouches, a sealed Tyvek-poly pouch within a foil-poly pouch. The foil-poly pouch is the sterile barrier for AVANCE. The immediate container closure system is packaged in a commercial carton, which is sealed with a Carton Seal Tape and placed in a zip top bag for freezer storage protection, see Figure 2.

Figure 2 (Image 02)

Figure 2 (Image 02)

Figure 2: AVANCE Packaging.
  • AVANCE is available in the following lengths and diameters (Table 4):
  • Table 4: Commercially Available Sizes of AVANCE
    Length (mm) Diameter (mm) NDC Numbers

    15

    1 to 2

    84545-115-01

    15

    2 to 3

    84545-215-01

    15

    3 to 4

    84545-315-01

    15

    4 to 5

    84545-415-01

    30

    1 to 2

    84545-130-01

    30

    2 to 3

    84545-230-01

    30

    3 to 4

    84545-330-01

    30

    4 to 5

    84545-430-01

    50

    1 to 2

    84545-150-01

    50

    2 to 3

    84545-250-01

    50

    3 to 4

    84545-350-01

    50

    4 to 5

    84545-450-01

    70

    1 to 2

    84545-170-01

    70

    2 to 3

    84545-270-01

    70

    3 to 4

    84545-370-01

    70

    4 to 5

    84545-470-01

16.2 Storage And Handling

  • Ensure the shipping container was received by the date indicated on the shipping container label and that sufficient dry ice is present within the shipping container. The amount of dry ice present within the shipping container can be assessed by the overall shipping container weight. The overall shipping container weight should be at least 12.0 lbs. for the 48-hour shipping container, 15.0 lbs. for the 72-hour shipping container, and 21.0 lbs. for the 120-hour shipping container.
  • Store AVANCE at ≤-40°C (≤-40°F) and keep frozen until use.
  • Do not use AVANCE if any component of the immediate container closure system or commercial carton has been compromised.
  • Thaw AVANCE to room temperature without heating using room temperature sterile normal saline or Lactated Ringer’s Solution (LRS). Do not implant a partially or fully frozen product. Once thawed, it should not be placed back into the freezer. Thawed AVANCE must be implanted or discarded within 12 hours.
  • Do not use if the lot number on the outer foil-poly pouch does not match the lot number on the outer carton.
  • Any expired or thawed AVANCE that is not used in a surgical procedure should be destroyed in accordance with local, state, and federal or country regulations for disposal of human tissue.
  • Contact Axogen Customer Care at 888-296-4361 for any issues with the condition of the packaging or graft, or for further instructions.

17 Patient Counseling Information

Discuss the following with the patient and/or caregivers.

  • Procedural Complications
    Inform patients that procedure-related complications such as infection, pain, swelling, bleeding, scarring, neuroma formation, or delayed wound healing may occur after nerve repair surgery using AVANCE. Advise that careful postoperative monitoring is important. [see Warnings and Precautions (5.1) and Adverse Reactions (6)].
  • Transmission of Infectious Diseases
    Inform patients that AVANCE is made from cadaveric human tissue and, despite donor screening and validated processing steps, the risk of transmitting infectious diseases cannot be completely eliminated. [see Warnings and Precautions (5.2)].
  • Advise the patient to closely follow the physician-prescribed rehabilitation program.

Principal Display Panel - 1-2 Mm X 15 Mm Ndc: 84545-115-01

Labels – Carton, Variable, And Side 1-2mm X 15mm (Image 04)

Labels – Carton, Variable, And Side 1-2mm X 15mm (Image 04)

Commercial Carton, Variable Product Identification Label, and Side Label, 1-2 mm x 15 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 2-3 Mm X 15 Mm Ndc: 84545-215-01

Labels – Carton, Variable, And Side 2-3mm X 15mm (Image 05)

Labels – Carton, Variable, And Side 2-3mm X 15mm (Image 05)

Commercial Carton, Variable Product Identification Label, and Side Label, 2-3 mm x 15 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 3-4 Mm X 15 Mm Ndc: 84545-315-01

Labels – Carton, Variable, And Side 3-4mm X 15mm (Image 06)

Labels – Carton, Variable, And Side 3-4mm X 15mm (Image 06)

Commercial Carton, Variable Product Identification Label, and Side Label, 3-4 mm x 15 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 4-5 Mm X 15 Mm Ndc: 84545-415-01

Labels – Carton, Variable, And Side 4-5mm X 15mm (Image 07)

Labels – Carton, Variable, And Side 4-5mm X 15mm (Image 07)

Commercial Carton, Variable Product Identification Label, and Side Label, 4-5 mm x 15 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 1-2 Mm X 30 Mm Ndc: 84545-130-01

Labels – Carton, Variable, And Side 1-2mm X 30mm (Image 08)

Labels – Carton, Variable, And Side 1-2mm X 30mm (Image 08)

Commercial Carton, Variable Product Identification Label, and Side Label, 1-2 mm x 30 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 2-3 Mm X 30 Mm Ndc: 84545-230-01

Labels – Carton, Variable, And Side 2-3mm X 30mm (Image 09)

Labels – Carton, Variable, And Side 2-3mm X 30mm (Image 09)

Commercial Carton, Variable Product Identification Label, and Side Label, 2-3 mm x 30 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 3-4 Mm X 30 Mm Ndc: 84545-330-01

Labels – Carton, Variable, And Side 3-4mm X 30mm (Image 10)

Labels – Carton, Variable, And Side 3-4mm X 30mm (Image 10)

Commercial Carton, Variable Product Identification Label, and Side Label, 3-4 mm x 30 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 4-5 Mm X 30 Mm Ndc: 84545-430-01

Labels – Carton, Variable, And Side 4-5mm X 30mm (Image 11)

Labels – Carton, Variable, And Side 4-5mm X 30mm (Image 11)

Commercial Carton, Variable Product Identification Label, and Side Label, 4-5 mm x 30 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 1-2 Mm X 50 Mm Ndc: 84545-150-01

Labels – Carton, Variable, And Side 1-2mm X 50mm (Image 12)

Labels – Carton, Variable, And Side 1-2mm X 50mm (Image 12)

Commercial Carton, Variable Product Identification Label, and Side Label, 1-2 mm x 50 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 2-3 Mm X 50 Mm Ndc: 84545-250-01

Labels – Carton, Variable, And Side 2-3mm X 50mm (Image 13)

Labels – Carton, Variable, And Side 2-3mm X 50mm (Image 13)

Commercial Carton, Variable Product Identification Label, and Side Label, 2-3 mm x 50 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 3-4 Mm X 50 Mm Ndc: 84545-350-01

Labels – Carton, Variable, And Side 3-4mm X 50mm (Image 14)

Labels – Carton, Variable, And Side 3-4mm X 50mm (Image 14)

Commercial Carton, Variable Product Identification Label, and Side Label, 3-4 mm x 50 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 4-5 Mm X 50 Mm Ndc: 84545-450-01

Labels – Carton, Variable, And Side 4-5mm X 50mm (Image 15)

Labels – Carton, Variable, And Side 4-5mm X 50mm (Image 15)

Commercial Carton, Variable Product Identification Label, and Side Label, 4-5 mm x 50 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel.-. 1-2 Mm X 70 Mm Ndc: 84545-170-01

Labels – Carton, Variable, And Side 1-2mm X 70mm (Image 16)

Labels – Carton, Variable, And Side 1-2mm X 70mm (Image 16)

Commercial Carton, Variable Product Identification Label, and Side Label, 1-2 mm x 70 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 2-3 Mm X 70 Mm Ndc: 84545-270-01

Labels – Carton, Variable, And Side 2-3mm X 70mm (Image 17)

Labels – Carton, Variable, And Side 2-3mm X 70mm (Image 17)

Commercial Carton, Variable Product Identification Label, and Side Label, 2-3 mm x 70 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 3-4 Mm X 70 Mm Ndc: 84545-370-01

Labels – Carton, Variable, And Side 3-4mm X 70mm (Image 18)

Labels – Carton, Variable, And Side 3-4mm X 70mm (Image 18)

Commercial Carton, Variable Product Identification Label, and Side Label, 3-4 mm x 70 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

Principal Display Panel - 4-5 Mm X 70 Mm Ndc: 84545-470-01

Labels – Carton, Variable, And Side 4-5mm X 70mm (Image 19)

Labels – Carton, Variable, And Side 4-5mm X 70mm (Image 19)

Commercial Carton, Variable Product Identification Label, and Side Label, 4-5 mm x 70 mm

Front Panel

AVANCE®
(acellular nerve allograft-arwx

Not representative of actual size,
visual appearance of graft may very

Adhere product side label here

Contains 1 sterile acellular nerve
allograft in a clamshell tray,
packaged in two sealed pouches

axogen®

* Please review the disclaimer below.