Warning: High Level Radiation Exposure With Use Of Incorrect Eluent And Failure To Follow The Eluate Testing Protocol
High Level Radiation Exposure with Use of Incorrect Eluent
Patients are exposed to high radiation levels when the CardioGen-82 generator is eluted with the incorrect eluent due to high strontium (Sr) 82 and Sr 85 breakthrough levels.
- Use only additive-free 0.9% Sodium Chloride Injection to elute the generator.
- If an incorrect solution is used to elute the CardioGen-82 generator:
- Strictly adhere to the generator eluate testing protocol, to minimize the risk of excess radiation exposure, including daily testing and additional testing at Alert Limits.
- Stop using the generator if it reaches any of its Expiration Limits:
Excess Radiation Exposure with Failure to Follow the Eluate Testing Protocol
Excess radiation exposure occurs when the levels of Sr 82 or Sr 85 in the Rubidium Chloride Rb 82 Injection exceed specified limits.